Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 20 de 38
Filter
1.
Clin Biochem ; 127-128: 110763, 2024 May.
Article in English | MEDLINE | ID: mdl-38615787

ABSTRACT

OBJECTIVES: Contamination with intravenous (IV) fluids is a common cause of specimen rejection or erroneous results in hospitalized patients. Identification of contaminated samples can be difficult. Common measures such as failed delta checks may not be adequately sensitive nor specific. This study aimed to determine detection criteria using commonly ordered tests to identify IV fluid contamination and validate the use of these criteria. METHODS: Confirmed contaminated and non-contaminated samples were used to identify patterns in laboratory results to develop criteria to detect IV fluid contamination. The proposed criteria were implemented at a tertiary care hospital laboratory to assess performance prospectively for 6 months, and applied to retrospective chemistry results from 3 hospitals and 1 community lab to determine feasibility and flagging rates. The algorithm was also tested at an external institution for transferability. RESULTS: The proposed algorithm had a positive predictive value of 92 %, negative predictive value of 91 % and overall agreement of 92 % when two or more criteria are met (n = 214). The flagging rates were 0.03 % to 0.07 % for hospital and 0.003 % for community laboratories. CONCLUSIONS: The proposed algorithm identified true contamination with low false flagging rates in tertiary care urban hospital laboratories. Retrospective and prospective analysis suggest the algorithm is suitable for implementation in clinical laboratories to identify samples with possible IV fluid contamination for further investigation.


Subject(s)
Algorithms , Humans , Retrospective Studies , Laboratories, Clinical , Prospective Studies , Specimen Handling/methods , Specimen Handling/standards
3.
Clin Biochem ; 125: 110731, 2024 Mar.
Article in English | MEDLINE | ID: mdl-38360198

ABSTRACT

BACKGROUND: An analytical benchmark for high-sensitivity cardiac troponin (hs-cTn) assays is to achieve a coefficient of variation (CV) of ≤ 10.0 % at the 99th percentile upper reference limit (URL) used for the diagnosis of myocardial infarction. Few prospective multicenter studies have evaluated assay imprecision and none have determined precision at the female URL which is lower than the male URL for all cardiac troponin assays. METHODS: Human serum and plasma matrix samples were constructed to yield hs-cTn concentrations near the female URLs for the Abbott, Beckman, Roche, and Siemens hs-cTn assays. These materials were sent (on dry ice) to 35 Canadian hospital laboratories (n = 64 instruments evaluated) participating in a larger clinical trial, with instructions for storage, handling, and monthly testing over one year. The mean concentration, standard deviation, and CV for each instrument type and an overall pooled CV for each manufacturer were calculated. RESULTS: The CVs for all individual instruments and overall were ≤ 10.0 % for two manufacturers (Abbott CVpooled = 6.3 % and Beckman CVpooled = 7.0 %). One of four Siemens Atellica instruments yielded a CV > 10.0 % (CVpooled = 7.7 %), whereas 15 of 41 Roche instruments yielded CVs > 10.0 % at the female URL of 9 ng/L used worldwide (6 cobas e411, 1 cobas e601, 4 cobas e602, and 4 cobas e801) (CVpooled = 11.7 %). Four Roche instruments also yielded CVs > 10.0 % near the female URL of 14 ng/L used in the United States (CVpooled = 8.5 %). CONCLUSIONS: The number of instruments achieving a CV ≤ 10.0 % at the female 99th-percentile URL varies by manufacturer and by instrument. Monitoring assay precision at the female URL is necessary for some assays to ensure optimal use of this threshold in clinical practice.


Subject(s)
Myocardial Infarction , Humans , Male , Female , Prospective Studies , Canada , Myocardial Infarction/diagnosis , Biological Assay , Troponin , Troponin T , Biomarkers , Reference Values
4.
Clin Biochem ; 116: 133-137, 2023 Jun.
Article in English | MEDLINE | ID: mdl-37146789

ABSTRACT

Sporadic mechanically-induced hemolysis associated with kinks in extracorporeal blood circuits during hemodialysis is a rare but potentially serious complication that exhibits laboratory features consistent with both in vivo and in vitro hemolysis. Misclassification of clinically significant hemolysis as in vitro can lead to inappropriate test cancellation and delayed medical interventions. Here, we report three cases of hemolysis attributed to kinked hemodialysis blood lines, which we have defined as "ex vivo" hemolysis. All three cases demonstrated an initial mixed picture of laboratory features consistent with both classifications of hemolysis. Specifically, absent features of in vivo hemolysis on blood film smear despite normal potassium led to the misclassification of these samples as in vitro hemolysis and their cancellation. A proposed mechanism for these overlapping laboratory features is the recirculation of damaged red blood cells from the kinked or pinched hemodialysis line back into the patient circulation producing an "ex vivo" hemolysis presentation. In two of the three cases, the patients developed acute pancreatitis as a result of hemolysis and required urgent medical follow up. We developed a decision pathway to help laboratories in identifying and handling these samples by recognizing that in vitro and in vivo hemolysis have overlapping laboratory features. These cases highlight the need for laboratorians and the clinical care team to be vigilant about mechanically-induced hemolysis from the extracorporeal circuit during hemodialysis. Communication is critical to identify the cause of hemolysis in these patients and prevent unnecessary delays in result reporting.


Subject(s)
Hemolysis , Pancreatitis , Humans , Acute Disease , Renal Dialysis/adverse effects , Erythrocytes
6.
CJC Open ; 5(12): 925-933, 2023 Dec.
Article in English | MEDLINE | ID: mdl-38204856

ABSTRACT

Background: The objective of this study was to assess the introduction of a high-sensitivity troponin I (hs-TnI) assay and its associated accelerated protocol on emergency department (ED) length of stay (LOS) for patients presenting with chest pain, compared to an accelerated diagnostic protocol using conventional troponin (TnI) testing. Methods: We conducted a retrospective cohort study of all adults with a primary presenting complaint of chest pain of cardiac origin and a Canadian Triage and Acuity Scale score of 2 or 3, between November 8, 2019 and November 9, 2021, to a tertiary-care urban Canadian ED. The primary outcome was ED LOS. Secondary outcomes included consultation proportions and major adverse cardiac events within 30 days of the index ED visit. Results: A total of 2640 patients presenting with chest pain were included, with 1333 in the TnI group and 1307 in the hs-TnI group. Median ED LOS decreased significantly, from 392 minutes for the TnI group, and 371 minutes for the hs-TnI group (median difference = 21 minutes; 95% confidence interval: 5.3, 36.7). The numbers of consultations and admissions were not statistically different between study periods. The major adverse cardiac events outcomes did not change following the implementation of the hs-TnI test (13.6% vs 13.1%; P = 0.71). Conclusions: The implementation of an accelerated chest pain protocol using an hs-TnI assay in a tertiary-care Canadian ED was associated with a modest reduction of LOS for all patients, and a substantial reduction of LOS for patients undergoing serial troponin testing. This strategy was safe, with no increase in adverse outcomes.


Contexte: Cette étude visait à évaluer l'introduction du dosage de la troponine I de haute sensibilité (hs-TnI) et le protocole accéléré qui lui est associé sur la durée des séjours aux urgences dans le cas des patients qui consultent pour une douleur thoracique, comparativement à un protocole diagnostique accéléré faisant appel à un test de troponine classique (TnI). Méthodologie: Nous avons mené une étude de cohorte rétrospective portant sur tous les adultes qui se sont présentés aux urgences d'un établissement urbain de soins tertiaires canadien entre le 8 novembre 2019 et le 9 novembre 2021 principalement pour une douleur thoracique d'origine cardiaque et dont le score était de 2 ou 3 à l'Échelle canadienne de triage et de gravité (ETG). Le principal critère d'évaluation était la durée du séjour au service des urgences. Les critères d'évaluation secondaires comprenaient la fréquence des consultations et les événements cardiaques indésirables majeurs dans les 30 jours ayant suivi la visite de référence aux urgences. Résultats: Au total, 2640 patients qui s'étaient présentés aux urgences pour une douleur thoracique ont été inclus, 1333 se trouvant dans le groupe TnI et 1307 dans le groupe hs-TnI. La durée médiane du séjour aux urgences a diminué considérablement, passant de 392 minutes dans le groupe TnI à 371 minutes dans le groupe hs-TnI (différence médiane de 21 minutes; intervalle de confiance [IC] à 95 % : 5,3-36,7). Les consultations et les admissions n'ont pas affiché de différence statistique entre les périodes de l'étude. Les événements cardiaques indésirables majeurs n'ont pas varié après l'introduction du dosage de la hs-TnI (13,6 % vs 13,1 %; p = 0,71). Conclusions: L'adoption d'un protocole accéléré pour la douleur thoracique à l'aide du dosage de la hs-TnI au service des urgences d'un établissement de soins tertiaires canadien a été associée à une légère réduction de la durée du séjour pour l'ensemble des patients et à une réduction substantielle de cette durée pour les patients soumis à des analyses de la troponine en série. De plus, cette stratégie était sûre sans hausse des événements indésirables.

7.
Clin Biochem ; 97: 62-66, 2021 Nov.
Article in English | MEDLINE | ID: mdl-34343576

ABSTRACT

OBJECTIVE: Reagent lot-to-lot comparisons are recommended by accreditation bodies to ensure that the performance of each reagent lot meets acceptable standards for quality patient results. The general approach is comprised of performing quality control (QC) and patient comparison between the old and new reagent lots and evaluating against a pre-defined criteria. Reagent lot comparison practices are often variable despite using the same instrument across different laboratories. This is costly, time consuming, and can lead to variability in acceptance criteria. While Clinical & Laboratory Standards Institute (CLSI) has a recommended guideline for reagent lot validation, it is often difficult to execute for small and rural laboratories due to limited resources. Defining the analytes required for detailed validation is important to allocate appropriate resources to ensure quality patient results. The goal of this study was to develop a standardized approach to reagent lot validation and optimize lab resources on Vitros chemistry instruments. DESIGN AND METHOD: This study consists of a retrospective and prospective analysis of reagent lot changes in dry slide chemistry analyzers (Ortho Clinical Diagnostics Vitros). Two years of retrospective reagent lot comparison data was obtained at a single site. A prospective study was conducted by assessing aliquots of 10 patient sample pools at 9 sites with Vitros analyzers. RESULTS: Of the 19 chemistry analytes evaluated, albumin, sodium, and total protein showed significant differences between reagent lots and also exceeded the pre-defined acceptance criteria. CONCLUSION: For these analytes, our recommendations are to perform a comprehensive lot validation with QC and patient samples. A simple lot validation with a reflex approach comprised of initially assaying QC can be adapted for the more stable analytes to allow achieving quality patient result in a resource constraint rural environment.


Subject(s)
Chemistry, Clinical/instrumentation , Chemistry, Clinical/standards , Reagent Kits, Diagnostic/standards , Equipment and Supplies , Humans , Prospective Studies , Quality Control , Reproducibility of Results , Retrospective Studies
8.
Clin Biochem ; 91: 31-38, 2021 May.
Article in English | MEDLINE | ID: mdl-33444605

ABSTRACT

OBJECTIVES: Hydroxocobalamin (OHCob) is an antidote for cyanide poisoning in patients rescued from house fires and is known to cause interference with certain laboratory tests. Consensus is lacking on the extent of this interference and on how to handle these samples. The objectives of this study were to characterize OHCob interference across a wide range of laboratory tests and to develop protocols for identifying and reporting these samples. DESIGNS & METHODS: Patient plasma samples (n = 5) were spiked with OHCob (1.5 mg/mL) and compared to controls without this drug. A series of analytes were measured using chemistry, urinalysis, coagulation, hematology, and blood gas instruments. Dose-response testing was performed on a subset of assays that showed interferences ≥10%. RESULTS: Of the 77 analytes evaluated, 27 (35%) showed interference from OHCob, with chemistry and coagulation analytes showing the greatest effects. Of those affected, 22 analytes had a positive interference, whereas 5 analytes had negative interference. Dose-response studies showed dose-dependent increases and/or decreases consistent with initial spiking studies. Although red in colour, plasma samples with OHCob did not trigger hemolysis index flags, necessitating a special sample identification and reporting protocol. CONCLUSION: OHCob had significant effects on several analytes across different instruments. These findings led to the development of special sample handling and reporting protocols to identify OHCob samples and ensure only accurate results are released. It is vital for emergency departments to document and notify their laboratories whenever blood samples from these patients are drawn.


Subject(s)
Antidotes/pharmacokinetics , Blood Chemical Analysis , Hydroxocobalamin/pharmacokinetics , Poisoning , Potassium Cyanide , Antidotes/administration & dosage , Female , Humans , Hydroxocobalamin/administration & dosage , Male , Poisoning/blood , Poisoning/drug therapy
9.
Clin Biochem ; 78: 25-31, 2020 Apr.
Article in English | MEDLINE | ID: mdl-31743687

ABSTRACT

OBJECTIVES: Previous analytical evaluations of the Beckman Coulter Access high sensitivity troponin (hsTn) I assay have focused on single platforms and laboratory sites. The purpose of this study was to determine assay robustness across different platforms at multiple sites, platform-specific characteristics, and equivalence to other hsTn methods in a large laboratory network. METHODS: Barricor plasma was used to assess imprecision, linearity, sensitivity (limit of blank and detection, LOB/LOD), and comparability to the conventional AccuTnI+3 and other hsTn assays. Various studies were conducted across a total of 9 laboratories using Beckman DxI800 and Access2 platforms. RESULTS: Within-laboratory precision was <10% across all target patient pool concentrations, however, DxI800 mean values were 20% higher than Access2 in the range of 3.6-44.9 ng/L. LOBs and LODs were lower on DxI800, 0.27 and 0.90 ng/L, respectively, compared to 2.9 and 3.2 ng/L, on Access2. Both showed excellent linearity across the full range. In method comparison to AccuTnI+3, DxI800 had a higher slope (0.9417 versus 0.8495) and positive bias (+18.1% versus -9.9%) compared to Access2, a trend further pronounced at concentrations <150 ng/L. At values <150 ng/L, there was good agreement with Abbott hsTnI (slope = 1.017, r = 0.932), but poor agreement with the Roche hsTnT assay (slope = 1.687, r = 0.589). Inter-laboratory split sample comparisons across 2 DxI800 and 7 Access2 sites showed close agreement, except at low concentrations <10 ng/L where DxI800 was 2.8 ng/L higher (p<0.001). CONCLUSIONS: The Beckman hsTnI assay showed robust analytical performance across different laboratories and platforms. However, discrepancies between platforms were found at low concentrations where rapid acute myocardial infarction (AMI) rule-out decisions occur. These differences have important implications for AMI risk assessment, suggesting that laboratories should develop platform-specific parameters rather than using them interchangibly.


Subject(s)
Blood Chemical Analysis/methods , Troponin I/blood , Biomarkers/blood , Female , Humans , Limit of Detection , Male , Sensitivity and Specificity
10.
J Appl Lab Med ; 3(5): 764-774, 2019 03.
Article in English | MEDLINE | ID: mdl-31639752

ABSTRACT

BACKGROUND: Cardiac troponin I (cTnI) 99th percentile cutoffs, used in the diagnosis of acute myocardial infarction, are not standardized across cTnI assays. We compared 3 point-of-care (POC) and 1 central laboratory contemporary cTnI assays against the Abbott high-sensitivity (hs) cTnI to evaluate the analytical concordance and the feasibility of using a single cutoff value for all assays. METHODS: Fresh blood samples collected from 102 inpatients in the coronary care unit were measured on central laboratory instruments (Beckman Coulter DxI AccuTnI+3 TnI, Abbott Architect hs-TnI) and cTnI POC analyzers (Alere Triage Troponin I, Radiometer AQT90, Abbott i-STAT). Agreement and correlation between the contemporary cTnI assays and hs-cTnI assay were assessed using regression analysis. Proportional bias was assessed using Bland-Altman plots. Concordance between the contemporary cTnI and hs-cTnI assays was determined by diagnostic contingency tables at specific cutoffs. RESULTS: Most POC cTnI assays had excellent correlation with the Abbott hs-cTnI method (r 2 = 0.955-0.970) except for Alere Triage (r 2 = 0.617), while proportional bias is evident between all cTnI assays. Overall concordance between POC contemporary cTnI assays and hs-cTnI assay was 80% to 90% at their respective 99th percentile cutoffs. The concordance increased to 90% to 95% when a fixed cutoff of 0.03 to 0.05 ng/mL was used across the assays. CONCLUSIONS: This study demonstrates poor analytical concordance between cTnI assays at the 99th percentile and supports the notion of a single clinical decision limit for cTnI and consequently standardization of diagnostic protocols despite the analytical differences among these assays.


Subject(s)
Biomarkers/blood , Laboratories/standards , Myocardial Infarction/blood , Myocardial Infarction/diagnosis , Point-of-Care Systems/statistics & numerical data , Troponin I/blood , Troponin T/blood , Biological Assay , Female , Humans , Male , Triage/statistics & numerical data
11.
Clin Chim Acta ; 496: 18-24, 2019 Sep.
Article in English | MEDLINE | ID: mdl-31201816

ABSTRACT

INTRODUCTION: The BD Barricor tube uses a novel mechanical separator designed to eliminate gel artifacts, decrease cellular contamination, and improve stability. Here, we evaluated the Barricor tube as a possible replacement for PST using Beckman Coulter analyzers under both optimal, alternative, and suboptimal centrifugation conditions based on BD recommendations. METHODS: Paired PST and Barricor samples were collected from 4 local hospitals and processed based on site-specific preanalytical systems involving automated or manual centrifugation. Centrifugation conditions ranged from 1912 ×g for 10 min (suboptimal), 2060 g for 10 min (alternative), and 4000 ×g for 3 or 10 min (optimal). Tube volume (4.5 vs. 5.5 ml) was also assessed. Forty-three chemistry and immunochemistry analytes were measured on Beckman Coulter DxC and DxI analyzers. RESULTS: Using an automated preanlaytical system with suboptimal spin conditions, no bias between PST and Barricor was observed for all analytes tested except lactate dehydrogenase (LD). Further investigation revealed significant increase in LD when Barricor was spun for 10 min at 1912, 2060 and 4000 ×g, ranging from +7.4-19.4% vs. PST across the entire measurement interval (87-493 U/l). Smaller tube volume was also associated with higher LD. Differences in LD occurred despite no change in other hemolysis markers such as potassium, phosphate, and AST. CONCLUSIONS: LD is most sensitive to varying centrifugation conditions (time and speed) in Barricor tubes. We recommend that BD centrifugation protocols should be closely evaluated to determine if Barricor is equivalent to PST under local preanalytical configurations.


Subject(s)
Blood Chemical Analysis , Blood Specimen Collection/instrumentation , Immunoassay , L-Lactate Dehydrogenase/blood , Plasma/chemistry , Humans
13.
Clin Biochem ; 50(15): 851-857, 2017 Oct.
Article in English | MEDLINE | ID: mdl-28583435

ABSTRACT

OBJECTIVES: BD Canada recently released a blood collection tube with a novel mechanical separator called the Barricor. We evaluated this tube as an alternate sample type for cardiac troponin I (cTnI) testing using the Beckman Coulter AccuTnI+3 assay. DESIGN AND METHODS: 3014 paired patient specimens (Barricor, plasma separator tube or PST) were obtained from the emergency departments and cardiac care units of nine hospitals in and around Edmonton, Alberta. After centrifugation, each plasma sample was analyzed for cTnI using the Beckman Coulter AccuTnI+3 assay. In addition, selected samples were analyzed multiple times within a single run or over 4-5days to generate imprecision data for the assay. RESULTS: Repeatability and within-laboratory studies revealed an imprecision of <10% at concentrations above 0.025µg/L for the Barricor as well as BD's traditional PST. Paired patient sample comparisons over the full range of the assay yielded linear regression slopes ranging from 0.956 to 1.011 and Pearson correlation coefficients ranging from 0.993 to 0.999. At a lower range of results closer to the manufacturer's 99th percentile cutoffs correlation was slightly worse, but still acceptable, with linear regression slopes ranging from 0.967 to 1.211 and Pearson correlation coefficients ranging from 0.983 to 0.987. Notably, at these lower concentrations the agreement between individual PST and Barricor results worsened with decreasing cTnI concentration. Differences between pairs of results became particularly large (-50 to +400%) at PST cTnI concentrations ≤0.015µg/L. Closer inspection of the data around the 0.02 and 0.04µg/L 99th percentile cutoffs revealed a number of discordances between PST and Barricor results, with at least some of these attributable to false elevations in the PST results. CONCLUSIONS: Together, our results suggest that the Barricor blood collection tube is good alternative to the traditional PST for cTnI testing using the AccuTnI+3 assay. The Barricor appears to minimize spurious, nonreproducible, and false elevations in cTnI results for a subset of patients but additional studies are needed to determine if it reduces overall false elevations. cTnI results below 0.04µg/L may still be of questionable accuracy even with the use of this new tube.


Subject(s)
Blood Chemical Analysis/instrumentation , Blood Chemical Analysis/methods , Blood Specimen Collection/instrumentation , Blood Specimen Collection/methods , Troponin I/blood , Female , Humans , Male , Reproducibility of Results
15.
Clin Chim Acta ; 468: 71-75, 2017 May.
Article in English | MEDLINE | ID: mdl-28131674

ABSTRACT

BACKGROUND: Milrinone is a potent selective phosphodiesterase type III inhibitor which stimulates myocardial function and improves myocardial relaxation. Although therapeutic monitoring is crucial to maintain therapeutic outcome, little data is available. A proof-of-principle study has been initiated in our institution to evaluate the clinical impact of optimizing milrinone dosing through therapeutic drug monitoring (TDM) in children following cardiac surgery. We developed a robust LC-MS/MS method to quantify milrinone in serum from pediatric patients in real-time. METHODS: A liquid-liquid extraction procedure was used to prepare samples for analysis prior to measurement by LC-MS/MS. Performance characteristics, such as linearity, limit of quantitation (LOQ) and precision, were assessed. Patient samples were acquired post-surgery and analyzed to determine the concentration-time profile of the drug as well as to track turn-around-times. RESULTS: Within day precision was <8.3% across 3 levels of QC. Between-day precision was <12%. The method was linear from 50 to 800µg/l; the lower limit of quantification was 22µg/l. Comparison with another LC-MS/MS method showed good agreement. Using this simplified method, turnaround times within 3-6h were achievable, and patient drug profiles demonstrated that some milrinone levels were either sub-therapeutic or in the toxic range, highlighting the importance for milrinone TDM. CONCLUSIONS: This simplified and quick method proved to be analytically robust and able to provide therapeutic monitoring of milrinone in real-time in patients post-cardiac surgery.


Subject(s)
Blood Chemical Analysis/methods , Chromatography, High Pressure Liquid/methods , Drug Monitoring/methods , Milrinone/blood , Tandem Mass Spectrometry/methods , Child , Humans , Liquid-Liquid Extraction , Milrinone/isolation & purification
18.
Clin Biochem ; 49(12): 879-84, 2016 Aug.
Article in English | MEDLINE | ID: mdl-27157715

ABSTRACT

OBJECTIVES: Studies have demonstrated improved analytical performance of the Nova StatStrip glucose meter, but limited data is available on its clinical performance in critically ill neonates in the neonatal intensive care unit (NICU). DESIGN AND METHODS: A retrospective charge review was conducted on 651 neonates admitted to the NICU over 2 years. Demographics, sample collection information, and clinical details were recorded. Glucose measurements were performed at the bedside using either the Nova StraStrip or LifeScan SureStep Flexx meters as well as corresponding measurements of laboratory venous plasma glucose. Performance was analyzed by receiver operator characteristic (ROC) curves for detecting hypoglycemia and critical glucose levels. RESULTS: Linear regression analysis comparing StatStrip and laboratory venous plasma glucose samples demonstrated significantly tighter agreement (r(2)=0.7994) and accuracy (mean bias=0.13mmol/L) than SureStep (r(2)=0.6845 and mean bias=0.53mmol/L). StatStrip also showed improved sensitivity for detecting critical low glucose values ≤3.0mmol/L (80.9 vs 68.9%, p<0.05). ROC curve analysis further demonstrated excellent performance of StatStrip at this cutoff with an AUC of 0.98. Overall, neonates were also tested significantly less frequently with the StatStrip meter by 24% compared to SureStep. CONCLUSIONS: Implementation of StatStrip led to better agreement with venous plasma glucose, improved detection of critical low glucose results, and more efficient test utilization. This study demonstrates the importance of accurate and sensitive glucose monitoring in the NICU.


Subject(s)
Biomarkers/blood , Blood Glucose/analysis , Hyperglycemia/diagnosis , Hypoglycemia/diagnosis , Intensive Care, Neonatal/standards , Point-of-Care Systems/standards , Reagent Strips , Female , Follow-Up Studies , Hematologic Tests , Humans , Hyperglycemia/blood , Hypoglycemia/blood , Infant, Newborn , Linear Models , Male , Monitoring, Physiologic/instrumentation , Practice Guidelines as Topic , Prognosis , ROC Curve , Reproducibility of Results , Retrospective Studies
19.
Clin Chim Acta ; 450: 196-202, 2015 Oct 23.
Article in English | MEDLINE | ID: mdl-26310965

ABSTRACT

BACKGROUND: The CALIPER program has previously reported a comprehensive database of pediatric reference intervals for 63 biochemical and immunochemical markers. Here, covariate-stratified reference intervals were determined for a number of special assays not previously reported. METHODS: A total of 1917 healthy children and adolescents were recruited and serum concentrations of 14 biochemical markers were measured using the Abbott Architect ci4100 system. Age and gender partitions were statistically determined, outliers removed and reference intervals calculated using CSLI C28-A3 guidelines. RESULTS: Many analytes showed dynamic changes in concentration requiring at least 3 age partitions. Unique intervals were required within the first year of life for: pancreatic amylase, C-peptide, ceruloplasmin, insulin, ß-2-microglobulin, cystatin C, dehydroepiandrosterone sulfate (DHEA-S), and α-1-glycoprotein. Cholinesterase, cholinesterase-dibucaine number, and immunoglobulin E required only 2 age partitions and α-1-antitrypsin required only one. Anti-CCP and anti-TPO levels were below the detection limit of the assay. Some analytes including insulin and DHEA-S required additional gender partitions for specific age groups. CONCLUSIONS: Complex profiles were observed for endocrine and special chemistry markers, requiring establishment of age- and gender-specific reference intervals. These updated reference intervals will allow improved laboratory assessment of pediatric patients but should be validated for each analytical platform and local population as recommended by CLSI.


Subject(s)
Biomarkers/blood , Blood Chemical Analysis/standards , Health , Residence Characteristics , Adolescent , Aging/blood , Child , Child, Preschool , Databases, Factual , Female , Growth and Development , Humans , Infant , Infant, Newborn , Male , Reference Values
SELECTION OF CITATIONS
SEARCH DETAIL
...